UI - Tesis Open :: Kembali

UI - Tesis Open :: Kembali

Pengukuran dan Simulasi Monte Carlo Persentase Dosis Kedalaman Berkas Sinar X 6 MV Jatuh pada Bidang Miring : Verifikasi TPS

Agung Nugroho Oktavianto; Djarwani Soeharso Soejoko, supervisor; Freddy Haryanto, supervisor; Mussadiq Musbach, examiner; Warsito, examiner; Setiawan Sutopo, examiner (Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2010)

 Abstrak

Telah dilakukan pengukuran relatif untuk mengetahui perubahan parameter dosimetri berkas sinar X 6 MV pada sumbu vertikal/sumbu normal terhadap permukaan dengan variasi sudut gantry 0º, 15º, 30º, 45º, dan 60º dari pesawat Linac Electa Precise 5991 milik Rumah Sakit Hasan Sadikin Bandung dengan luas lapangan 10x10 cm2, 15x15cm2 , dan 20x20 cm2. Rekonstruksi TPS dan simulasi Monte Carlo menggunakan parameter yang sama dengan pengukuran.
Hasil penelitian menunjukkan bahwa terjadi perubahan kedalaman maksimum yang dipengaruhi kenaikan luas lapangan dan sudut gantri dengan deviasi terbesar -33,3% pada luas lapangan 20x20 cm2 dan sudut gantri 600 terhadap luas lapangan 10x10 cm2 dan sudut gantri 00. Terjadi titik belok pada sumbu vertikal yang disebabkan karena adanya kemiringan permukaan, yang menurun dengan kenaikan sudut gantri. Hasil rekonstruksi TPS menunjukkan kesesuaian terhadap hasil pengukuran pada titik PDD maksimum, kedalaman 5 cm dan 10 cm, berbeda dengan hasil perhitungan Monte Carlo yang relatif lebih tinggi. Demikian pula untuk titik-titik di luar sumbu vertikal/sumbu normal.
Rekonstruksi TPS mempunyai deviasi < 2%, sedangkan perhitungan Monte Carlo mempunyai deviasi < 2% hanya pada daerah kuadran (+). Pengamatan titik-titik pada sumbu utama berkas hanya dilakukan dengan rekonstruksi TPS karena keterbatasan kemampuan alat, hasil penelitian menunjukkan bahwa PDD pada kedalaman dmax, 5 cm, dan 10 cm pada sumbu utama untuk sudut gantri 150 sampai dengan 600 cenderung menurun untuk ukuran lapangan yang sama demikian juga pada daerah kuadran (+).

Relative meauserement for change of dosimetry parameters X ray beam 6 MV Linac Electa Precise 5991 property of Hasan Sadikin Bandung?s hospital at the vertical central line from surface with incident obliquity 0º, 15º, 30º, 45º, 60º and 10x10 cm2, 15x15cm2, 20x20 cm2 field has been done. As in Reconstruction of TPS and Monte Carlo simuation with the same parameters measurement have been conducted.
The result of the experiment showed the change of maximum depth because of the increase in the filed dan angle of gantri with deviation -33% at the 20x20 cm2 field and 600 angle of gantry to 10x10 cm2 field and 00 angle of gantry . To happened point of to drop at the vertical central line because change sloping field to surface and drop on with increase of angle?s gantry.
The result of reconstruction of TPS to show not different with meaurement at point máximum PDD, 5 cm depth, 10 cm depth, point at off verical line, Monte Carlo calculation is higher. In all oint of reconstruction?s TPS has deviation < 2%, but calculation Monte Carlo just happen at the kuadran (+) area. Observation at the point in the central beam just done with reconstruction of TPS because the equipment capability limitation in the measurement, PDD at the depth of dmax, 5 cm, 10 cm, 150 to 600 angle of gantry in the central beam tilted decrease for the same field and kuadran (+) area too.

 File Digital: 1

 Metadata

Jenis Koleksi : UI - Tesis Open
No. Panggil : T29077
Entri utama-Nama orang :
Entri tambahan-Nama orang :
Program Studi :
Subjek :
Penerbitan : Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2010
Bahasa : ind
Sumber Pengatalogan : LibUI ind rda
Tipe Konten : text
Tipe Media : unmediated ; computer
Tipe Carrier : volume ; online resource
Deskripsi Fisik : xii, 115 pages : illustration ; 29 cm + appendix
Naskah Ringkas :
Lembaga Pemilik : Universitas Indonesia
Lokasi : Perpustakaan UI, Lantai 3
  • Ketersediaan
  • Ulasan
  • Sampul
No. Panggil No. Barkod Ketersediaan
T29077 15-19-620012650 TERSEDIA
Ulasan:
Tidak ada ulasan pada koleksi ini: 20279762
Cover